Can a phone software update change speech biomarker readings?

Yes. An iOS or Android update, new phone firmware, or a new release of the recording app or browser can change how the phone filters, levels and compresses the microphone signal before analysis, so speech biomarker readings can move while the participant’s speech has not changed. In Cephalgo’s own checks for its CTAD 2026 poster P446 (doi:10.5281/zenodo.24297337), a minor iOS release changed audio processing in measurement mode on both iPhones checked.

Explainer for trial sponsors and digital endpoint scientists, by Cephalgo · Updated November 2026

Audio chain

Where a phone processes speech before it is saved

Between the microphone and the saved file, a phone passes the signal through firmware, the operating system’s audio processing, the recording mode the app asks for and often a compression codec. Each layer is software that an update can change.

Voice processing

Noise suppression, echo cancellation and automatic gain control are built into both platforms, mainly for calls. On iPhone, voice chat mode pairs with a voice‑processing input whose gain control is on by default. On Android, the platform can insert these effects by default, depending on the audio source.

Recording modes

Apps ask for a mode, not for untouched audio. Apple’s measurement mode is the iPhone recording mode for apps that measure sound and need to minimise system‑supplied processing; it disables some dynamics processing (the automatic levelling of loud and soft sounds), not all of it. Android’s UNPROCESSED source falls back to the default source on phones that do not support it.

Microphone modes

On iPhone, the user picks Standard, Voice Isolation or Wide Spectrum in Control Center; Voice Isolation attenuates everything but the voice. An app can read the mode but cannot set it, and the active mode can differ when the audio route does not support it.

App, browser and file

The app sets the sample rate and whether audio is saved compressed. In a browser, echo cancellation, noise suppression and gain control are W3C getUserMedia constraints.

The audio chain in a phoneEach layer is software
  1. Participant’s voice
  2. Microphone and firmwareCan change
  3. Operating system voice processingCan change
  4. Recording mode the app requestsCan change
  5. App or browser settingsCan change
  6. Codec and file formatCan change
  7. Saved audio file

Android’s rules for VOICE_RECOGNITION switch off noise reduction and gain control by default, not compressors or limiters, and phones need not offer the unprocessed source. In the stock firmware configurations Cephalgo read, the voice‑recognition path of many Android phones is configured with a compressor, and so is the default recording mode of every iPhone model read. What phone firmware does to speech ›

What updates change

What a software update can change

An update can change any layer it ships: the tuning inside a recording mode, which processing runs by default, the codec, or the settings an app or browser requests. Updates install on the phone’s own schedule, not the study’s. A shift in speech measures caused this way, sometimes called drift, is a change in the audio, not in the speaker.

  1. Tuning inside a mode

    A new build can retune the filters, noise suppression or gain stages behind a recording mode while the mode keeps its name. Apple documents what measurement mode is for, not a fixed signal chain.

  2. Default processing

    On Android, which effects run by default is set per device, and apps are told to check the default on each session. Phone makers set that chain in firmware and can change it in a new build.

  3. Codecs

    Android’s software media codecs are a Mainline module that Google Play system updates can update. Audio compressed on the phone can change with a codec update; uncompressed audio cannot.

  4. App releases

    A new app release can switch the audio source, mode, sample rate or format. Android phones must offer 16‑bit capture at five sample rates from 8 to 48 kHz (CDD 5.4.1), so the rate is the app’s choice.

  5. Browser defaults

    The W3C specification leaves the default for echo cancellation to each browser and notes that browsers tend to turn it on, a default chosen to suit real‑time calls.

How often

How often phones change during a trial

Often. An iPhone 11 or later kept up to date received fifteen iOS updates between 20 August 2025 (iOS 18.6.2) and 17 August 2026 (iOS 26.6.1), according to Apple’s security releases page; iOS 27 followed in September 2026.

iPhone

Of those fifteen updates on Apple’s security releases page, one was a major release (iOS 26), six were minor releases and eight were point releases.

Android

Android Security Bulletins are published on the first Monday of each month. Pixel phones get a monthly update with security patches and functional improvements, and other makers release their own builds.

Apps and browsers

Capture apps, eCOA apps and browsers release on their own schedules.

Updates do not arrive evenly across sites, countries or arms, so a change can add variance, or a step a longitudinal model reads as decline or improvement. How phone changes can look like decline or a treatment effect ›

Published research

What published studies show

When the same voices are recorded or passed through different devices, apps and processing, fundamental frequency usually holds, while spectral, cepstral, perturbation, intensity and timing measures shift, sometimes systematically.

Software can shift measures

Sanker et al., 2021 compared recording equipment and remote recording software with solid‑state audio. Measures of frequency, duration and noise moved, and “some measurements are systematically shifted in particular recording conditions.”

Platform enhancements

Weerathunge et al., 2021 sent 29 voice samples from people with dysphonia through six video‑conferencing conditions. Every measure except mean fundamental frequency changed significantly, most with one platform’s enhancements switched on.

Bias by measure

Jannetts et al., 2019 recorded 22 speakers on a studio microphone and several smartphones. Cepstral peak prominence showed significant bias, and the random error of jitter and shimmer was judged problematic.

Different but correlated

Awan et al., 2024 compared headset microphones and two smartphones with a research reference on 24 vowel samples. Some recording configurations differed significantly from the reference despite strong correlations (rs > .90); the two smartphones used directly showed no significant differences for the measures assessed.

Compression

Oreskovic et al., 2024 converted 17,298 smartphone recordings from 505 participants to three compressed formats. Effects depended on format, tool, bitrate and feature. Zhang et al., 2021 found lossless phone recordings closer to a field recorder than a video‑meeting app for formants.

These studies compare devices, platforms or formats side by side; none of the studies above follows one phone through an operating system update. Cephalgo’s posters simulate that effect by holding the recordings fixed and changing only the processing.

What Cephalgo found

What Cephalgo found

In Cephalgo’s own checks for its CTAD 2026 poster P446 (Boston, 16–19 November 2026; doi:10.5281/zenodo.24297337), a minor iOS release changed audio processing in measurement mode on both iPhones checked, and Android and web browser updates also changed audio processing settings.

  1. After the simulated update

    After the worst simulated audio processing update, the sensitivity and specificity of Cephalgo’s multi-stream model, which combines linguistic, acoustic and self-supervised analysis, were 17 to 19 percentage points higher than with standard handcrafted features (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).

  2. One update moved people across groups

    In the same Cephalgo poster, on three published speech datasets, one simulated audio processing update moved 32% of people to the other group in the hardest comparison, Alzheimer’s disease versus MCI in one dataset, with standard handcrafted speech features and 27% with Cephalgo’s multi-stream model, although their condition had not changed (doi:10.5281/zenodo.24297337).

  3. A false trend, mostly removed by recalibration (simulated)

    In Cephalgo’s simulated year-long trials with a known decline added, one audio processing update made speech timing features decline up to 39% faster or slower than the true decline, and recalibration from the audio removed 95% to 97% of that false trend (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).

  4. Same recordings, different noise reduction

    In Cephalgo’s ISCTM 2026 poster (Philadelphia, 16 September 2026; doi:10.5281/zenodo.24060676), recordings from the Italian Parkinson’s Voice and Speech corpus (28 people with Parkinson’s, 22 age-matched controls; 136 recordings from 9 unseen participants) went through three noise-reduction methods standing in for successive phone software versions, and the choice of noise-reduction method explained 2% of the variance in the multi-stream model’s output, against 22% for a handcrafted-feature model. Metrics come from those 9 held-out participants (5 with Parkinson’s, 4 controls; 72 and 64 recordings). Under the change of noise-reduction method, the multi-stream model’s worst-case AUC was 0.927 and its per-recording ICC(A,1) 0.751; the limits are one controlled cohort and a simulated change, and the worst-case specificity of 60.9% rests on 4 control speakers. AUC measures how well a model separates the two groups; AUC runs from 0 to 1, where 0.5 is chance and 1 is perfect separation. ICC(A,1) measures how closely each recording’s reading agrees when only the noise-reduction method changes (1 means identical).

  5. Six update scenarios

    In Cephalgo’s poster at the Digital Biomarkers in Clinical Trials Summit (Basel, 23 June 2026; doi:10.5281/zenodo.24060707), on a public Parkinson’s speech dataset (n=50: 28 patients, 22 controls), four noise-cancellation algorithms across six scenarios simulated a mid-trial update; Cephalgo’s pipeline kept sensitivity between 86.4% and 97.7% in all six, while three comparator architectures (HuBERT-base, Wav2Vec2-base, gradient-boosted handcrafted features) fell as low as 51–56%.

Both Parkinson’s results above use the same public corpus. The recordings come from the Italian Parkinson’s Voice and Speech corpus (Dimauro and Girardi, IEEE DataPort, CC BY 4.0); in Cephalgo’s posters they were passed through noise-reduction methods, and the dataset’s authors have not reviewed these results.

What regulators ask

What the FDA asks about phone software updates

The FDA’s December 2023 final guidance on digital health technologies names operating system updates directly: record each update, make sure measurements are unaffected and, if they changed, compare data before and after and say how differences are handled before unblinding.

“If a DHT or associated technology, such as a general computing platform, is updated during a clinical investigation (e.g., operating system update), sponsors should ensure that the DHT remains fit‑for‑purpose, such that the updates do not affect the measurements and that verification and validation studies (see section IV.C of this guidance) are still applicable.” “Significant changes in the measurement after updates may invalidate the results from a clinical investigation.”
U.S. Food and Drug Administration, Digital Health Technologies for Remote Data Acquisition in Clinical Investigations, final guidance, December 2023, section IV.H.4, DHT updates and other changes

The same section asks sponsors to “keep a record of the timing and nature of any updates for each DHT”, for sponsor‑provided and participants’ own phones alike.

What to do

What sponsors can do

Treat the phone’s software as part of the measurement: log it, test measures against processing changes and fix the before‑and‑after comparison in the analysis plan.

  1. Log versions per recording

    Record the device model and software version with every recording: operating system build, app or browser version, recording and microphone mode, input route and analysis version.

  2. Ask for the least‑processed path

    Measurement mode on iPhone; UNPROCESSED on Android where declared, otherwise VOICE_RECOGNITION; in browsers, processing constraints set to false and read back. None is guaranteed to stay fixed across updates.

  3. Keep audio uncompressed

    Save uncompressed or lossless files at capture, so a codec update cannot change them.

  4. Test on the same audio

    Run the trial’s speech measures on the same recordings processed in different ways, before the first participant and when a new version enters the study.

  5. Ask the vendor

    Ask how its speech measures behave when audio processing changes, and for the evidence.

  6. Plan before unblinding

    State in the analysis plan how readings from before and after a change are compared.

The full set of controls, from study phones to recalibration: How do you keep speech readings stable when phones and software change? ›

Questions

Phone software updates and speech readings: common questions

Can an iPhone software update change speech biomarker readings in a clinical trial?

Yes. In Cephalgo’s own checks, a minor iOS release changed audio processing in measurement mode on both iPhones checked (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337). Measurement mode is the iPhone mode meant to minimise system processing, so even the most neutral path can change.

Does an iOS or Android update change microphone audio processing?

It can. Noise suppression, echo cancellation, gain control and the tuning of each recording mode live in the phone’s software and firmware. In Cephalgo’s own checks, a minor iOS release changed processing in measurement mode, and Android and web browser updates also changed audio processing settings.

Which recording settings give the least processed speech audio on a phone?

On iPhone, measurement mode. On Android, the UNPROCESSED source where the phone declares support, otherwise VOICE_RECOGNITION, which starts with noise reduction and gain control off. In a browser, echoCancellation, noiseSuppression and autoGainControl set to false. None is guaranteed to stay fixed across updates.

How often do phones update during a clinical trial?

An iPhone 11 or later kept up to date received fifteen iOS updates between 20 August 2025 (iOS 18.6.2) and 17 August 2026 (iOS 26.6.1), according to Apple’s security releases page; iOS 27 followed in September 2026. Android security bulletins are published every month, and apps and browsers update on their own schedules.

What does the FDA say about operating system updates during a trial?

Its December 2023 guidance asks sponsors to record the timing and nature of updates, make sure measurements are unaffected, compare data before and after if they changed, and specify before unblinding how differences will be addressed.

Can speech readings stay stable when phone software changes?

Yes. Cephalgo’s readings are recalibrated from the recorded audio. In simulated year-long trials, recalibration from the audio removed 95% to 97% of the false trend that one audio processing update created in speech timing features (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).

Do voice isolation and automatic gain control change voice features in speech recordings?

They can. Voice Isolation, which an iPhone user can switch on in Control Center, attenuates everything but the voice. Automatic gain control adjusts the microphone level automatically, so measured loudness can change with it; it is on by default on the iPhone voice-processing input, and on some Android phones the platform inserts it by default, depending on the audio source. An app can read the iPhone microphone mode but cannot set it, so log it with each recording. In Cephalgo’s own checks, a minor iOS release changed audio processing even in measurement mode, the mode meant to minimise it (doi:10.5281/zenodo.24297337).

Cephalgo

About Cephalgo

Cephalgo SAS, founded in 2021 in Strasbourg, France, builds reliable speech readings for CNS clinical trials in Alzheimer’s disease, Parkinson’s disease, Huntington’s disease and ALS. Cephalgo’s readings are recalibrated from the recorded audio to stay stable when participants’ own phones and software change. Device changes ›BYOD ›Study phones ›

Funded by the European Union. Cephalgo’s work described on this page was supported by the European Innovation Council (EIC) Accelerator (project ADAPTE, Grant Agreement No. 190129251) and by Horizon Europe (grant agreement No 101095436). Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union, the European Innovation Council and SMEs Executive Agency (EISMEA) or the European Health and Digital Executive Agency (HaDEA). Neither the European Union nor the granting authorities can be held responsible for them.

Sources

Guidance, platform documentation and studies

Every platform statement on this page was checked against the official documentation on 27 November 2026.

Regulator

Platform documentation

  • Apple Developer Documentation. AVAudioSession.Mode.measurement and AVAudioSession.Mode.voiceChat.

    Measurement mode is for apps that need to minimise system-supplied signal processing and disables some dynamics processing; voice chat mode tunes the input for voice, and the voice-processing input it pairs with applies echo cancellation and automatic gain control.

  • Apple Developer Documentation. AVCaptureDevice.MicrophoneMode and isVoiceProcessingAGCEnabled.

    Standard, Voice Isolation and Wide Spectrum microphone modes, chosen by the user in Control Center; automatic gain control on the voice-processing input is enabled by default.

  • Apple. Apple security releases. Accessed 27 November 2026.

    Release dates of every iOS update, used here to count the updates between August 2025 and August 2026.

  • Android Developers. MediaRecorder.AudioSource and NoiseSuppressor (with AutomaticGainControl and AcousticEchoCanceler).

    Audio sources tuned for voice recognition, voice communication or unprocessed sound; the platform can insert noise suppression, gain control or echo cancellation by default depending on the source.

  • Android Open Source Project. Android 17 Compatibility Definition Document, sections 5.4.1, 5.4.2 and 5.11.

    Required capture formats and sample rates, noise reduction and gain control off by default for voice recognition, and only a level multiplier allowed on the unprocessed source, which phones need not offer.

  • Android Open Source Project. Android Security Bulletins, Pixel Update Bulletins and Mainline (modular system components).

    Bulletins published on the first Monday of each month; monthly Pixel updates with security patches and functional improvements; software media codecs ship as a module that Google Play system updates can update.

  • W3C. Media Capture and Streams. Candidate Recommendation Draft, 19 November 2026, sections 4.3.9 and 11.

    Defines the echoCancellation, autoGainControl and noiseSuppression constraints for getUserMedia; the browser sets the default for echo cancellation.

Published studies

  • Sanker C, Babinski S, Burns R, et al. (Don’t) try this at home! The effects of recording devices and software on phonetic analysis. Language. 2021;97(4):e360–e382. doi:10.1353/lan.2021.0075

    Recording equipment and software had large effects on frequency, duration and noise measures, some of them systematic shifts.

  • Weerathunge HR, Segina RK, Tracy L, Stepp CE. Accuracy of acoustic measures of voice via telepractice videoconferencing platforms. Journal of Speech, Language, and Hearing Research. 2021;64(7):2586–2599. doi:10.1044/2021_JSLHR-20-00625

    Transmission condition affected every acoustic measure except mean fundamental frequency; one platform with enhancements switched on had the largest effects.

  • Jannetts S, Schaeffler F, Beck J, Cowen S. Assessing voice health using smartphones: bias and random error of acoustic voice parameters captured by different smartphone types. International Journal of Language & Communication Disorders. 2019;54(2):292–305. doi:10.1111/1460-6984.12457

    Cepstral peak prominence showed systematic bias across smartphones; the random error of jitter and shimmer was judged problematic.

  • Awan SN, Bahr R, Watts S, Boyer M, Budinsky R, Bridge2AI Voice Consortium, Bensoussan Y. Validity of acoustic measures obtained using various recording methods including smartphones with and without headset microphones. Journal of Speech, Language, and Hearing Research. 2024;67(6):1712–1730. doi:10.1044/2024_JSLHR-23-00759

    Some recording methods, including smartphones with a low-cost headset microphone, gave significantly different acoustic measures that still correlated above 0.90 with a research-grade reference; smartphones used directly did not differ significantly.

  • Oreskovic J, Kaufman J, Fossat Y. Impact of audio data compression on feature extraction for vocal biomarker detection: validation study. JMIR Biomedical Engineering. 2024;9:e56246. doi:10.2196/56246

    Compression effects on voice features depended on the format, the conversion tool, the bitrate and the feature.

  • Zhang C, Jepson K, Lohfink G, Arvaniti A. Comparing acoustic analyses of speech data collected remotely. Journal of the Acoustical Society of America. 2021;149(6):3910–3916. doi:10.1121/10.0005132

    Lossless phone recording apps matched a field recorder more closely than a video-meeting app, which showed unexpected intensity drops.

Dataset

  • Dimauro G, Girardi F. Italian Parkinson’s Voice and Speech. IEEE DataPort, 2019. doi:10.21227/aw6b-tg17. Licence CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/).

    Source of the Parkinson’s and control recordings in the ISCTM and Basel posters; in those posters the recordings were passed through noise-reduction methods, and the dataset’s authors have not reviewed the results.

Cephalgo posters

  • Noise‑cancellation variability as a source of longitudinal drift in speech‑based digital endpoints for Alzheimer’s disease trials. CTAD 2026, Boston, 16–19 November 2026, poster P446. doi:10.5281/zenodo.24297337

    Device checks, group changes after a simulated audio processing update, and recalibration in simulated year-long trials.

  • When the device changes mid‑trial: speech‑based digital endpoint robustness to on‑device preprocessing variability. ISCTM 2026 Autumn Conference, Philadelphia, 16 September 2026. doi:10.5281/zenodo.24060676

    The same Parkinson’s and control recordings run through three noise-reduction methods.

  • When the device changes mid‑trial: evaluating speech‑based digital endpoint robustness to on‑device preprocessing variability in longitudinal remote monitoring. Digital Biomarkers in Clinical Trials Summit, Basel, 23 June 2026. doi:10.5281/zenodo.24060707

    Four noise-cancellation algorithms across six scenarios simulating a mid-trial update.

Which phones and software versions does your protocol allow?

Tell us, and we will send you the full CTAD 2026 poster P446 on phone and software updates. If your study is already recording, a stability audit checks each reading across those changes and recalibrates it from the audio.

Request

We use your work email to send you the document and to follow up about it and related work. You can opt out at any time. Privacy policy

Thank you. We will send it to your work email.

Something went wrong. Please write to contact@cephalgo.com.

Thank you! Your submission has been received!
Oops! Something went wrong while submitting the form.