What are the risks of recording speech on participants’ own phones (BYOD) in clinical trials?

With bring your own device (BYOD), the phone becomes part of the speech measurement: each participant records through a different microphone and audio processing chain, and software updates or a new phone can change it mid‑study, so readings can differ or shift for reasons unrelated to health. Cephalgo SAS builds reliable speech readings that are recalibrated from the recorded audio, so they stay stable when participants’ own phones and software change (doi:10.5281/zenodo.24297337).

Explainer for trial sponsors and digital endpoint scientists, by Cephalgo · Updated November 2026

Benefits

Why trials record speech on participants’ own phones

Own phones give more readings between visits with less burden: participants use a phone they already carry, and the sponsor has no study phones to manage.

More readings between visits

In mPower, a study run entirely through an iPhone app, 5,826 participants made 65,022 voice recordings (Bot et al., 2016).

Lower burden

The FDA notes that familiar technology may reduce the burden of extra devices, and it may also support compliance (Demanuele et al., 2022). Representatives of the Critical Path Institute’s ePRO Consortium called provisioned devices costly for sponsors and burdensome to carry (Gwaltney et al., 2015).

Why speech is different from a questionnaire

For on‑screen questionnaires, a review found best‑practice design enough to conclude measurement equivalence, including for BYOD, given a minimum device specification and standard response scales (Byrom et al., 2019). In a speech task, the microphone and the phone’s audio processing produce the data, so the device is part of the measurement.

Six risks

Six measurement risks of BYOD speech recording

The risks come from one fact: in BYOD the recording chain is different for each participant, and it can change during the study without anyone being told.

A different chain for every participant

Microphones, recording modes and built‑in processing differ by model. Android’s unprocessed microphone source falls back to the default path where a phone does not support it. On iPhone, the participant can pick Voice Isolation in Control Center, which attenuates sounds other than the voice. In the stock firmware configurations Cephalgo read, the Snapdragon Galaxy S24 Ultra’s voice-recognition path is configured with a compressor that narrows the loudness range; the Exynos Galaxy S24 configuration has none. Noise reduction and speech measures ›

Updates on each person’s schedule

An iPhone 11 or later kept up to date received fifteen iOS updates between 20 August 2025 (iOS 18.6.2) and 17 August 2026 (iOS 26.6.1), according to Apple’s security releases page; iOS 27 followed in September 2026. Apple’s 9 November 2026 releases, iOS 27.2.1 and iOS 26.7.3, cover the same iPhone models, so one model can run either version. Software updates and speech readings ›

A new phone mid‑study

Phones are lost, broken and upgraded, and a replacement brings a new microphone and new processing: a step change for that participant. The FDA lists replacement and upgrades among the problems a risk management plan should address.

A device mix that differs by site, country or arm

Phone ownership can differ between sites and countries, and each arm’s mix changes as people update or replace phones. How phone changes can look like decline ›

Missing data and compatibility

Older operating systems lose support, the app may not install on some phones, and some candidates have no suitable phone. The FDA notes that remote collection may change the type and amount of missing data.

Room and distance

Each participant records in a different room, at a distance and angle nobody controls. Room acoustics and background noise affected voice parameters more than the type of microphone (Bottalico et al., 2020), and handheld phone distance and positioning may be highly variable (Awan et al., 2024).

Phone studies

What studies comparing phones found

Smartphones can capture usable voice recordings, but different phones do not give the same values, and how far they disagree depends on the measure.

Uloza et al., 2015

Sustained /a/ from 118 people, 84 with pathological voices, was recorded at once on a studio microphone and a Samsung Galaxy Note3. Measures correlated strongly (r = 0.73 to 1.0), and the phone recordings separated normal from pathological voices.

Grillo et al., 2016

Voice measures were captured with smartphones and a head‑mounted microphone and computed with three analysis programs; device and software choice made no significant difference to within-speaker measures.

Jannetts et al., 2019

22 speakers recorded on a studio microphone and several smartphones. Cepstral peak prominence, a measure of how clearly periodic the voice is, showed significant bias, and the random error of jitter and shimmer (cycle-to-cycle variation in pitch and in loudness) was judged problematic.

Awan et al., 2024

24 vowel samples played through a head‑and‑thorax model were recorded on reference equipment, headset microphones and two phones (iPhone 13 Pro, Google Pixel 6). Some recording configurations differed significantly from the reference despite strong correlations (rs > .90); the two smartphones used directly, at 2.5 cm, showed no significant differences for the measures assessed.

What this means for a BYOD trial

Grillo’s finding of no within-speaker difference sits beside the systematic differences that Jannetts found across smartphones and Awan found with some headset set-ups: agreement shown for one set of phones and measures need not hold for another. A strong correlation means phones rank people alike, not that they give the same value. Most of these studies also recorded under controlled conditions that a BYOD trial does not have. Android and iPhone recordings compared ›

Confounding

When the phone becomes a confounder

If the phone mix differs between the groups compared, a model can partly learn the phone instead of the condition, and a group difference can partly reflect devices.

A probe, a model given only the metadata and no audio, can test this. AUC measures how well a score separates two groups; 1 is perfect separation. In Cephalgo’s summary of results on the mPower study (doi:10.5281/zenodo.24583552), sustained vowels recorded unsupervised on participants’ own iPhones were analysed on 80 held-out participants (40 reporting Parkinson’s, 40 controls); a probe using only age, sex, phone model and app version reached AUC 0.61 for Parkinson’s versus controls, where 0.5 is chance, so details other than the voice carried some group information. Parkinson’s status was self-reported by app users; this is a research result, not a diagnostic or clinical decision tool. These data were contributed by users of the Parkinson mPower mobile application as part of the mPower study developed by Sage Bionetworks and described in Synapse (https://www.synapse.org/mPower).

Randomisation balances arms on average, but small arms can still end up with different phone and software mixes. The FDA guidance states: “The same method for collection of data should be used in all study arms.”

What regulators expect

What the FDA and EMA say about own phones

Both accept participant-owned phones and expect the sponsor to show that measurements stay consistent across the models and software versions it accepts.

The verification and validation process may include “Confirmation, either by new studies or by reference to existing data, that measurements across protocol‑specified DHTs are consistent when the protocol permits use of more than one brand or model of DHT to collect the same data in a clinical investigation.”
U.S. Food and Drug Administration, Digital Health Technologies for Remote Data Acquisition in Clinical Investigations, final guidance, December 2023, section IV.C.1

FDA: updates on own phones too

Sponsors should plan for updates whether devices are sponsor-provided or the participant’s own, record the timing and nature of each update and, if measurements changed, compare data before and after (section IV.H.4).

EMA: every accepted model and version

EMA’s guideline on computerised systems and electronic data in clinical trials (2023, section A5.1.3) asks for minimum device specifications, data quality and integrity across all accepted models and versions, an alternative for participants without a suitable device, and an application not exclusive to one model or operating system. It notes that the sponsor “has no control over the implementation of updates to the operating system”.

On FDA updates: Do phone and software changes shift speech endpoints in clinical trials? ›

What Cephalgo found

What Cephalgo found

In Cephalgo’s posters, audio processing changes moved standard speech measures more than Cephalgo’s multi-stream model.

  1. Updates changed audio processing

    In Cephalgo’s own checks for its CTAD 2026 poster P446 (Boston, 16–19 November 2026; doi:10.5281/zenodo.24297337), a minor iOS release changed audio processing in measurement mode on both iPhones checked, and Android and web browser updates also changed audio processing settings. Measurement mode is Apple’s recording mode for minimal system processing.

  2. People moved group with no change in condition

    In the same Cephalgo poster, on three published speech datasets, one simulated audio processing update moved 32% of people to the other group in the hardest comparison, Alzheimer’s disease versus MCI in one dataset, with standard handcrafted speech features and 27% with Cephalgo’s multi-stream model, although their condition had not changed (doi:10.5281/zenodo.24297337). Handcrafted features are measures each computed by a fixed formula; Cephalgo’s multi-stream design combines linguistic, acoustic and self-supervised analysis of the same recording.

  3. Changing only the noise reduction

    In Cephalgo’s ISCTM 2026 poster (Philadelphia, 16 September 2026; doi:10.5281/zenodo.24060676), recordings from the Italian Parkinson’s Voice and Speech corpus (28 people with Parkinson’s, 22 age-matched controls; 136 recordings from 9 unseen participants) went through three noise-reduction methods standing in for successive phone software versions, and the choice of noise-reduction method explained 2% of the variance in the multi-stream model’s output, against 22% for a handcrafted-feature model. Metrics come from those 9 held-out participants (5 with Parkinson’s, 4 controls; 72 and 64 recordings). Under the change of noise-reduction method, the multi-stream model’s worst-case AUC was 0.927 and its per-recording ICC(A,1) 0.751; the limits are one controlled cohort and a simulated change, and the worst-case specificity of 60.9% rests on 4 control speakers. ICC(A,1) scores agreement between repeated readings of the same recording, where 1 means exact agreement; specificity is the share of controls correctly identified. The recordings come from the Italian Parkinson’s Voice and Speech corpus (Dimauro and Girardi, IEEE DataPort, CC BY 4.0); in Cephalgo’s posters they were passed through noise-reduction methods, and the dataset’s authors have not reviewed these results.

How Cephalgo recalibrates readings from the audio ›

What to do

How to manage BYOD risks in a speech study

Keep the benefits of own phones by treating the phone as part of the measurement: specify it, log it, check every recording, test stability, recalibrate from the recorded audio, and plan the analysis before unblinding.

  1. Set a minimum phone specification

    Operating system versions the maker still supports and a recording check at enrolment, without limiting entry to the newest models.

  2. Fix the recording path

    Request the least‑processed recording mode each phone offers, log which one the app got, and store audio uncompressed so a codec update cannot change it.

  3. Log versions with every recording

    Phone model, operating system build, app version, microphone or headset in use, and analysis version, so every recording can be placed in time.

  4. Check every recording

    Read noise, clipping and speech duration from the audio, ask for a retake when a check fails, and give plain instructions on room and phone position.

  5. Watch the device mix

    Tabulate phone model and software version by site and visit, test whether a metadata-only probe predicts diagnosis or baseline severity, and treat operating system, version and model as possible confounders (Demanuele et al., 2022).

  6. Keep a study‑phone option

    Offer a provisioned phone to people without a suitable one. Study phones ›

Stability tests, recalibration and the analysis plan: How do you keep speech readings stable when phones and software change? ›BYOD with Cephalgo ›

Questions

BYOD speech recording: common questions

What are the risks of BYOD for speech recording in clinical trials?

Each participant records through a different microphone and audio processing chain, updates and new phones change that chain mid-study, the phone mix can differ between sites and arms, some phones lose support, and rooms and distances vary. Speech measures can then differ or shift for reasons unrelated to health.

Which speech biomarker gives reliable readings when participants use their own phones?

Ask for three checks: each measure tested on the same audio processed in different ways, as different phones and updates do; the phone model and software version logged with every recording; and a quality check on every recording. Cephalgo’s readings are recalibrated from the recorded audio to stay stable when participants’ own phones and software change; in its ISCTM 2026 poster, the choice of noise-reduction method explained 2% of the variance in the multi-stream model’s output, against 22% for a handcrafted-feature model (doi:10.5281/zenodo.24060676).

How much do speech measures vary between phones?

It depends on the measure. Studies that recorded the same voices on several phones found strong correlations with studio microphones, but also systematic bias for some measures, such as cepstral peak prominence, and random error judged problematic for others, such as jitter and shimmer.

Can the phone model confound a BYOD speech study?

Yes, if the mix of phones differs between the groups compared. A model can partly learn the phone instead of the condition, so check the phone mix by site and arm and test whether a probe given only device metadata predicts diagnosis.

Does the FDA allow participants to use their own phones in clinical trials?

Yes. The FDA’s December 2023 guidance on digital health technologies asks sponsors to weigh the advantages and disadvantages of participant-owned technology, says validation may include confirming that measurements are consistent across the brands and models allowed, and asks sponsors to plan for updates and to offer sponsor-provided devices so that no one is excluded.

About Cephalgo

About Cephalgo

Cephalgo SAS, founded in 2021 in Strasbourg, France, builds reliable speech readings for CNS clinical trials in Alzheimer's disease, Parkinson's disease, Huntington's disease and ALS. Cephalgo's readings are recalibrated from the recorded audio to stay stable when participants' own phones and software change. It reads speech from participants’ own phones, study phones, site‑visit recordings and audio a trial already holds. BYOD ›Study phones ›Device changes ›

Funded by the European Union. Cephalgo's work described on this page was supported by the European Innovation Council (EIC) Accelerator (project ADAPTE, Grant Agreement No. 190129251) and by Horizon Europe (grant agreement No 101095436). Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union, the European Innovation Council and SMEs Executive Agency (EISMEA) or the European Health and Digital Executive Agency (HaDEA). Neither the European Union nor the granting authorities can be held responsible for them.

Sources

Guidance, studies and records

The guidance, studies and records behind this page, each opened and checked on 27 November 2026.

Regulators

BYOD in clinical research

  • Gwaltney C, Coons SJ, O’Donohoe P, O’Gorman H, Denomey M, Howry C, Ross J. “Bring Your Own Device” (BYOD): The Future of Field‑Based Patient‑Reported Outcome Data Collection in Clinical Trials? Therapeutic Innovation & Regulatory Science. 2015;49(6):783–791. doi:10.1177/2168479015609104

    ePRO Consortium review of operational, privacy and regulatory considerations; provisioned devices are costly and burdensome.

  • Byrom B, Gwaltney C, Slagle A, Gnanasakthy A, Muehlhausen W. Measurement Equivalence of Patient‑Reported Outcome Measures Migrated to Electronic Formats: A Review of Evidence and Recommendations for Clinical Trials and Bring Your Own Device. Therapeutic Innovation & Regulatory Science. 2019;53(4):426–430. doi:10.1177/2168479018793369

    Best‑practice design is enough to conclude equivalence of on‑screen questionnaires, including for BYOD, given a minimum device specification and standard response scales.

  • Demanuele C, Lokker C, Jhaveri K, Georgiev P, Sezgin E, Geoghegan C, Zou KH, Izmailova E, McCarthy M. Considerations for Conducting Bring Your Own “Device” (BYOD) Clinical Studies. Digital Biomarkers. 2022;6(2):47–60. doi:10.1159/000525080

    BYOD data are more heterogeneous; device model, operating system and version as possible confounders; distribution across arms.

  • Bot BM, et al. The mPower study, Parkinson disease mobile data collected using ResearchKit. Scientific Data. 2016;3:160011. doi:10.1038/sdata.2016.11

    A Parkinson’s study run through an iPhone app; 9,520 participants shared data, 5,826 made 65,022 voice recordings.

Studies of phones, microphones and rooms

  • Uloza V, Padervinskis E, Vegiene A, et al. Exploring the feasibility of smart phone microphone for measurement of acoustic voice parameters and voice pathology screening. European Archives of Oto‑Rhino‑Laryngology. 2015;272(11):3391–3399. doi:10.1007/s00405-015-3708-4

    Studio microphone and smartphone recorded at once; strong correlations (r = 0.73 to 1.0).

  • Grillo EU, Brosious JN, Sorrell SL, Anand S. Influence of Smartphones and Software on Acoustic Voice Measures. International Journal of Telerehabilitation. 2016;8(2):9–14. doi:10.5195/ijt.2016.6202

    Device and software choice made no significant difference to within-speaker voice measures.

  • Jannetts S, Schaeffler F, Beck J, Cowen S. Assessing voice health using smartphones: bias and random error of acoustic voice parameters captured by different smartphone types. International Journal of Language & Communication Disorders. 2019;54(2):292–305. doi:10.1111/1460-6984.12457

    Bias for cepstral peak prominence; jitter and shimmer random error judged problematic.

  • Awan SN, Bahr R, Watts S, Boyer M, Budinsky R, Bridge2AI Voice Consortium, Bensoussan Y. Validity of Acoustic Measures Obtained Using Various Recording Methods Including Smartphones With and Without Headset Microphones. Journal of Speech, Language, and Hearing Research. 2024;67(6):1712–1730. doi:10.1044/2024_JSLHR-23-00759

    Significant but predictable differences for some headset set-ups; smartphones used directly did not differ significantly; handheld distance and positioning vary; headset recommended.

  • Bottalico P, Codino J, Cantor‑Cutiva LC, et al. Reproducibility of Voice Parameters: The Effect of Room Acoustics and Microphones. Journal of Voice. 2020;34(3):320–334. doi:10.1016/j.jvoice.2018.10.016

    Room acoustics and background noise affected voice parameters more than the microphone.

Platform documentation

  • Apple. Apple security releases. Accessed 27 November 2026.

    iOS release dates and the devices each release covers.

  • Apple Developer Documentation. AVCaptureDevice.MicrophoneMode and preferredMicrophoneMode.

    Standard, Voice Isolation and Wide Spectrum modes; the user selects the mode in Control Center.

  • Android Developers. MediaRecorder.AudioSource.

    The unprocessed source is used if available and behaves like the default source otherwise.

Datasets

Cephalgo records

Own phones or study phones: which is your team weighing?

Tell us what you measure and on which phones, and we will send you the full CTAD 2026 poster P446. For a protocol in design, a prospective study gives site-visit and home readings per participant, pre-specified in your statistical analysis plan.

Request

We use your work email to send you the document and to follow up about it and related work. You can opt out at any time. Privacy policy

Thank you. We will send it to your work email.

Something went wrong. Please write to contact@cephalgo.com.