How do you keep speech readings stable when phones and software change?
Fix and log the recording chain at capture, test each speech measure on the same audio processed different ways, and recalibrate readings from the recorded audio, with the method fixed in the analysis plan before unblinding. Cephalgo SAS recalibrates its speech readings from the recorded audio; in its CTAD 2026 poster P446 (doi:10.5281/zenodo.24297337), this removed 95% to 97% of a false trend that one audio processing update created in simulated year-long trials. Sponsors set capture and logging; testing and recalibration are done with the speech analysis provider, and the checklist shows who does each step.
Explainer for trial sponsors and digital endpoint scientists, by Cephalgo · Updated November 2026
Capture: choose, fix and log the audio path
Stability starts at capture: the least-processed mode on each platform, lossless audio at a fixed sample rate, and versions logged with every file, so any later change can be dated.
iPhone
Apple’s measurement mode, a recording setting, is meant to minimise the signal processing iOS adds to the sound and records from the primary microphone. Participants can also pick a microphone mode, such as voice isolation, in Control Center; the app can read the active mode, so log it.
Android
The unprocessed source allows nothing beyond a level multiplier (a plain volume scaling), but only on phones that declare it. Otherwise, voice recognition must have noise reduction and gain control off by default (Android 17 Compatibility Definition, 5.11 and 5.4.2).
Web browser
A web app can ask the browser to turn off noise suppression, echo cancellation and automatic gain control (W3C Media Capture), then read back what was applied with getSettings() and store it.
Lossless files
When 17,298 smartphone recordings were converted to MP3, M4A or WMA, the effect on features depended on format, tool and feature (Oreskovic et al., 2024). Store audio uncompressed or lossless.
A fixed sample rate
Set one sample rate and log it with each file. Android requires its voice recognition source to capture at 44.1 or 48 kHz.
Versions on every recording
Log device model, operating system build, app version, analysis version, recording mode and input route with each file. The FDA asks for a record of the timing and nature of every update.
Study phones
Device management slows updates without stopping them: Apple’s declarative management defers iOS updates by at most 90 days, and Android allows freeze periods of at most 90 days, at least 60 days apart. Study phones ›
Own phones
Updates follow each participant. An iPhone 11 or later kept up to date received fifteen iOS updates between 20 August 2025 (iOS 18.6.2) and 17 August 2026 (iOS 26.6.1), according to Apple’s security releases page; iOS 27 followed in September 2026. The FDA asks sponsors to weigh this choice and to offer a provided device to anyone without one. BYOD ›
Validation: test each measure on the same audio, processed different ways
Use fixed-recording processing tests as one part of validation. Add controlled recordings through the actual supported capture configurations, with acceptance criteria for measurement agreement and preservation of the endpoint’s relevant clinical changes.
Start from the target population
Real recordings of patients and controls doing the trial’s tasks, from participants not used in training.
Process them the ways phones do
Noise reduction such as spectral subtraction, Wiener filtering or wavelet denoising, lossy codecs at several bit rates, and gain and level control, each standing in for a phone or software version.
Measure agreement
Report the intraclass correlation coefficient (ICC), which shows how closely repeated readings of the same speaker agree, in its absolute-agreement form ICC(A,1), and Bland–Altman bias with 95% limits of agreement. A consistency ICC ignores a uniform shift; Koo and Li (2016) advise absolute agreement for test–retest designs, and Bland and Altman (1986) showed that correlation misleads about agreement.
Check group separation
Report AUC (the area under the ROC curve, where 0.5 is chance and 1 is perfect separation), sensitivity and specificity, or the link to the clinical scale, in each condition. In Awan et al. (2024), some recording configurations differed significantly from the reference despite strong correlations (rs > .90); the two smartphones used directly showed no significant differences for the measures assessed.
Part of analytical validation
In the DiMe V3 framework, analytical validation tests the algorithms that turn sensor data into metrics (Goldsack et al., 2020). The FDA says that where an update may affect measurements, it may be necessary to validate them again, for example with previously collected data.
Repeat when something new enters
Run the panel again for each new phone model, operating system or app version, before unblinding.
Analysis: recalibrate, harmonise and pre‑specify
Readings from different processing periods are put back on one scale by recalibration from the recorded audio or by harmonisation, with covariates and sensitivity analyses fixed before unblinding.
“If changes to the measurement have occurred after the update, sponsors should compare data from DHTs before and after the update. Sensitivity analyses may be necessary to evaluate the impact of the update. […] Sponsors should specify how these differences will be addressed in the analysis of the trial prior to unblinding (if applicable).”
Recalibration from the recorded audio
The processing period comes from the device and software version logged with each recording or, where none was logged, is read from the audio; the files stay untouched. Cephalgo’s method ›
Harmonisation from imaging
ComBat, a statistical adjustment for batch effects first built for gene expression data (Johnson et al., 2007), removed scanner differences in diffusion MRI while preserving age effects (Fortin et al., 2017) and harmonised cortical thickness across 11 scanners (Fortin et al., 2018).
Longitudinal harmonisation
In 663 ADNI participants, longitudinal ComBat found additive and multiplicative scanner effects and, in simulations, controlled type I error (false positive findings) better than a scanner covariate (Beer et al., 2020).
Device and version as covariates
Device, operating system and app version can be covariates, but participant and device details can carry group information; check how devices spread across arms and sites first. In Cephalgo’s summary of results on the mPower study (doi:10.5281/zenodo.24583552), sustained vowels recorded unsupervised on participants’ own iPhones were analysed on 80 held-out participants (40 reporting Parkinson’s, 40 controls); a probe using only age, sex, phone model and app version reached AUC 0.61 for Parkinson’s versus controls, where 0.5 is chance, so details other than the voice carried some group information. Parkinson’s status was self-reported by app users; this is a research result, not a diagnostic or clinical decision tool. These data were contributed by users of the Parkinson mPower mobile application as part of the mPower study developed by Sage Bionetworks and described in Synapse (https://www.synapse.org/mPower).
In the analysis plan, before unblinding
Write in the version log, the method, the covariates, and sensitivity analyses such as pre‑change recordings only, estimates per period, and results with and without recalibration. Lock the adjustment and the analysis version before unblinding, fit it blind to treatment arm, and check that it keeps clinical change. Speech measures in the SAP ›How phone changes can look like decline ›
Questions to ask your speech analysis provider
Ask any provider for evidence, not assurances, of how its measures behave when the audio path changes. How to choose a provider ›
Which recording mode, sample rate and file format were your measures validated on?
Expect the mode on each platform, and lossless files.
What is logged with every recording?
The full version list above, each entry dated.
How does each measure behave on the same audio processed differently?
The validation panel above, on held-out participants.
How do you find a change nobody logged?
Ask whether the change can be read from the audio itself.
What happens to readings after a change?
Ask whether they are recalibrated, with definitions fixed, audio untouched and treatment arm blinded.
What will you write into our analysis plan?
Draft text on versions, covariates and sensitivity analyses, ready before unblinding.
What Cephalgo found
Cephalgo checked real updates, then changed only the audio processing and followed what happened to speech readings.
Updates change the processing
In Cephalgo’s own checks for its CTAD 2026 poster P446 (Boston, 16–19 November 2026; doi:10.5281/zenodo.24297337), a minor iOS release changed audio processing in measurement mode on both iPhones checked, and Android and web browser updates also changed audio processing settings.
After the simulated update
After the worst simulated audio processing update, the sensitivity and specificity of Cephalgo’s multi-stream model, which combines linguistic, acoustic and self-supervised analysis, were 17 to 19 percentage points higher than with standard handcrafted features (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).
People moved to the other group
In the same Cephalgo poster, on three published speech datasets, one simulated audio processing update moved 32% of people to the other group in the hardest comparison, Alzheimer’s disease versus MCI in one dataset, with standard handcrafted speech features and 27% with Cephalgo’s multi-stream model, although their condition had not changed (doi:10.5281/zenodo.24297337).
Recalibration removed most of the simulated false trend
In Cephalgo’s simulated year-long trials with a known decline added, one audio processing update made speech timing features decline up to 39% faster or slower than the true decline, and recalibration from the audio removed 95% to 97% of that false trend (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).
A denoiser change in Parkinson’s speech
In Cephalgo’s ISCTM 2026 poster (Philadelphia, 16 September 2026; doi:10.5281/zenodo.24060676), recordings from the Italian Parkinson’s Voice and Speech corpus (28 people with Parkinson’s, 22 age-matched controls; 136 recordings from 9 unseen participants) went through three noise-reduction methods standing in for successive phone software versions, and the choice of noise-reduction method explained 2% of the variance in the multi-stream model’s output, against 22% for a handcrafted-feature model. Under the change of noise-reduction method, the multi-stream model’s worst-case AUC was 0.927 and its per-recording ICC(A,1) 0.751; the limits are one controlled cohort and a simulated change, and the worst-case specificity of 60.9% rests on 4 control speakers.
Six deployment scenarios
In Cephalgo’s poster at the Digital Biomarkers in Clinical Trials Summit (Basel, 23 June 2026; doi:10.5281/zenodo.24060707), on a public Parkinson’s speech dataset (n=50: 28 patients, 22 controls), four noise-cancellation algorithms across six scenarios simulated a mid-trial update; Cephalgo’s pipeline kept sensitivity between 86.4% and 97.7% in all six, while three comparator architectures (HuBERT-base, Wav2Vec2-base, gradient-boosted handcrafted features) fell as low as 51–56%. The recordings come from the Italian Parkinson’s Voice and Speech corpus (Dimauro and Girardi, IEEE DataPort, CC BY 4.0); in Cephalgo’s posters they were passed through noise-reduction methods, and the dataset’s authors have not reviewed these results.
How Cephalgo keeps readings reliable when phones change
In Cephalgo’s method design, each recording is placed in the period in which one processing setting was in force, from the device and software version logged with it or, where none was logged, read from the audio, and the components of the multi-stream reading that are most stable across processing states anchor the estimate of each measure’s shift between periods. Measure definitions stay fixed, and shifts are delivered as recalibrated values or a covariate. Recalibration is applied to the measures and runs blind to treatment arm; the audio files are never altered. Device changes ›Software updates and speech readings ›
Checklist: who does what at each stage
The sponsor can set the first list; the second is done with the speech analysis provider.
- Capture
- Least‑processed recording mode named for each platform
- Capture
- Audio stored uncompressed or lossless, at a fixed sample rate
- Capture
- Device, operating system build, app version and recording mode logged per file
- Phones
- Study phones or own phones chosen; study‑phone updates deferred, timed and logged
- Analysis plan
- Version handling, covariates and sensitivity analyses written before unblinding
- Validation
- Each measure tested on the same audio processed different ways
- Validation
- ICC(A,1), Bland–Altman bias and group separation reported per condition
- Analysis
- Analysis version pinned and logged with every reading
- Analysis
- Readings recalibrated from the recorded audio, blind to treatment arm
- Vendor
- Written answers to the six questions above
Keeping speech readings stable: common questions
How do you recalibrate a speech biomarker after a phone switch?
Cephalgo places each recording in the processing period it was made in, then re-estimates each measure's shift between periods from the recorded audio, with measure definitions fixed, the audio left untouched and treatment arm blinded. In Cephalgo's simulated year-long trials, recalibration from the audio removed 95% to 97% of the false trend that one audio processing update created in speech timing features (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).
How do you harmonise speech features across devices in a multi-site study?
Log the recording mode per platform, test each feature on the same audio processed different ways, and put readings on one scale, either by harmonisation methods such as longitudinal ComBat, used in multi-site imaging and needing checks on speech data, or by recalibration from the recorded audio, which is what Cephalgo does for its readings. In Cephalgo's ISCTM 2026 poster, the choice of noise-reduction method explained 2% of the variance in the multi-stream model's output, against 22% for a handcrafted-feature model (doi:10.5281/zenodo.24060676).
How do you keep a speech endpoint stable across phone models in a multi-year trial?
Use the least-processed recording mode, lossless audio and version logging; choose measures that keep absolute agreement and group separation on differently processed audio; and pre-specify recalibration and sensitivity analyses in the analysis plan. Cephalgo's readings are recalibrated from the recorded audio to stay stable when participants' own phones and software change.
What does the FDA DHT guidance say about a device change mid-study?
The FDA's December 2023 guidance asks sponsors to record the timing and nature of updates, compare data before and after an update that changed measurements, consider sensitivity analyses, and specify before unblinding how differences will be addressed.
Do study phones remove the need for recalibration?
No. Study phones reduce variation between participants, but they still update, so each update needs checking and readings are recalibrated where it moves them. Device management can hold iOS and Android updates back for at most 90 days, and a replacement phone can arrive with newer software.
About Cephalgo
Cephalgo SAS, founded in 2021 in Strasbourg, France, builds reliable speech readings for CNS clinical trials in Alzheimer’s disease, Parkinson’s disease, Huntington’s disease and ALS. Cephalgo’s readings are recalibrated from the recorded audio to stay stable when participants’ own phones and software change. Device changes ›Validation approach ›
More in this series
Phone changes and speech endpoints · Software updates and speech readings · Risks of own phones (BYOD) · Noise reduction and speech measures · Phone changes as false decline · Android and iPhone recordings
Funded by the European Union. Cephalgo’s work described on this page was supported by the European Innovation Council (EIC) Accelerator (project ADAPTE, Grant Agreement No. 190129251) and by Horizon Europe (grant agreement No 101095436). Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union, the European Innovation Council and SMEs Executive Agency (EISMEA) or the European Health and Digital Executive Agency (HaDEA). Neither the European Union nor the granting authorities can be held responsible for them.
Guidance, papers and platform documentation
Every external factual statement on this page comes from these documents, each opened and checked on 27 November 2026.
Guidance and frameworks
U.S. Food and Drug Administration. Digital Health Technologies for Remote Data Acquisition in Clinical Investigations. Final guidance, December 2023. Sections IV.A.2, IV.A.4, IV.C.1 and IV.H.4.
Asks sponsors to name models and versions, weigh own devices against provided ones, log every update, compare data before and after a change, and say how differences are handled before unblinding.
Goldsack JC, et al. Verification, analytical validation, and clinical validation (V3): the foundation of determining fit‑for‑purpose for Biometric Monitoring Technologies (BioMeTs). npj Digital Medicine. 2020;3:55. doi:10.1038/s41746-020-0260-4
The DiMe V3 framework; analytical validation evaluates the algorithms that turn sensor data into metrics.
Agreement and harmonisation
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. 1986;327(8476):307–310. doi:10.1016/S0140-6736(86)90837-8
Showed that correlation misleads about agreement and set out the difference‑against‑mean analysis with limits of agreement.
Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. Journal of Chiropractic Medicine. 2016;15(2):155–163. doi:10.1016/j.jcm.2016.02.012
Explains the ten ICC forms, the difference between consistency and absolute agreement, and advises absolute agreement for test–retest.
Johnson WE, Li C, Rabinovic A. Adjusting batch effects in microarray expression data using empirical Bayes methods. Biostatistics. 2007;8(1):118–127. doi:10.1093/biostatistics/kxj037
Introduced ComBat, an empirical Bayes adjustment for batch effects that holds up in small batches.
Fortin JP, et al. Harmonization of multi‑site diffusion tensor imaging data. NeuroImage. 2017;161:149–170. doi:10.1016/j.neuroimage.2017.08.047
In 205 healthy participants on two scanners, ComBat removed site variation in diffusion measures while preserving age effects.
Fortin JP, et al. Harmonization of cortical thickness measurements across scanners and sites. NeuroImage. 2018;167:104–120. doi:10.1016/j.neuroimage.2017.11.024
Across 11 scanners in two multi‑site studies, ComBat removed scanner variability and increased statistical power.
Beer JC, et al. Longitudinal ComBat: A method for harmonizing longitudinal multi‑scanner imaging data. NeuroImage. 2020;220:117129. doi:10.1016/j.neuroimage.2020.117129
In 663 ADNI participants, found additive and multiplicative scanner effects; longitudinal ComBat controlled type I error better than a scanner covariate.
Speech and smartphone audio
Oreskovic J, Kaufman J, Fossat Y. Impact of audio data compression on feature extraction for vocal biomarker detection: validation study. JMIR Biomedical Engineering. 2024;9:e56246. doi:10.2196/56246
Converted 17,298 smartphone recordings to MP3, M4A and WMA; the effect on voice features depended on format, tool and feature.
Awan SN, et al. Validity of acoustic measures obtained using various recording methods including smartphones with and without headset microphones. Journal of Speech, Language, and Hearing Research. 2024;67(6):1712–1730. doi:10.1044/2024_JSLHR-23-00759
Some recording methods, including smartphones with a low-cost headset microphone, gave significantly different acoustic measures that still correlated above 0.90 with a research-grade reference; smartphones used directly did not differ significantly.
Bot BM, et al. The mPower study, Parkinson disease mobile data collected using ResearchKit. Scientific Data. 2016;3:160011. doi:10.1038/sdata.2016.11
An observational Parkinson’s study run entirely through an iPhone app, with repeated sensor recordings from participants with and without Parkinson’s.
Dimauro G, Girardi F. Italian Parkinson’s Voice and Speech. IEEE DataPort, 2019. doi:10.21227/aw6b-tg17. Licence CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/).
The Parkinson’s and control recordings used in Cephalgo’s ISCTM and Basel posters, where they were passed through noise-reduction methods.
Platform documentation
Apple. AVAudioSession.Mode.measurement; AVCaptureDevice.MicrophoneMode. Apple Developer Documentation.
Measurement mode minimises system‑supplied signal processing; microphone modes include voice isolation, chosen by the user in Control Center.
Apple. SoftwareUpdateSettingsDeferralsObject. Apple Developer Documentation; Apple security releases, Apple Support.
Managed iOS devices can defer operating system updates by 1 to 90 days; the release list dates every iOS update.
Android Open Source Project. MediaRecorder.AudioSource; SystemUpdatePolicy. Android Developers reference.
The unprocessed source behaves like the default source where it is not available; managed devices can freeze updates for at most 90 days at a time.
Android Open Source Project. Android 17 Compatibility Definition, sections 5.4.2 and 5.11.
Voice recognition capture must have noise reduction and gain control off by default; unprocessed capture allows only a level multiplier.
W3C. Media Capture and Streams. Candidate Recommendation Draft, 19 November 2026.
Defines the echoCancellation, autoGainControl and noiseSuppression settings a web app can control and read back.
Cephalgo posters and records
Noise‑cancellation variability as a source of longitudinal drift in speech‑based digital endpoints for Alzheimer’s disease trials. doi:10.5281/zenodo.24297337
Audio processing updates on phones and browsers; group changes and false trends after a simulated update; recalibration from the audio.
When the device changes mid‑trial: speech‑based digital endpoint robustness to on‑device preprocessing variability. doi:10.5281/zenodo.24060676
Parkinson’s and control recordings under three noise‑reduction methods, with agreement and group separation per model.
When the device changes mid‑trial: evaluating speech‑based digital endpoint robustness to on‑device preprocessing variability in longitudinal remote monitoring. doi:10.5281/zenodo.24060707
Four noise‑cancellation algorithms across six deployment scenarios simulating a mid‑trial update.
Smartphone speech and the patient‑rated MDS‑UPDRS speech item in Parkinson’s disease: a multi‑stream model evaluated on mPower data. doi:10.5281/zenodo.24583552
Includes a probe showing that age, sex, phone model and app version alone carried some group information.