How can phone and software changes look like decline or a treatment effect in a trial?

A phone or software update can shift a speech measure in one step. If the analysis does not account for that step, a longitudinal model can read it as faster decline, slower decline or a drug effect. In Cephalgo's simulated year-long trials with a known decline added, one audio processing update made speech timing features decline up to 39% faster or slower than the true decline, and recalibration from the audio removed 95% to 97% of that false trend (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).

Explainer for trial sponsors and digital endpoint scientists, by Cephalgo · Updated November 2026

Step or decline

A step in the measurement, not a change in the person

A phone or software update changes the measurement from the day it installs, in every recording that follows. The timing alone does not prove the cause, because a participant's speech can also change suddenly.

Updates are frequent

An iPhone 11 or later kept up to date received fifteen iOS updates between 20 August 2025 (iOS 18.6.2) and 17 August 2026 (iOS 26.6.1), according to Apple's security releases page; iOS 27 followed in September 2026.

Imaging met this first

In 208 adults scanned twice about two years apart on two MRI scanners of the same model, scanner changes over time could cancel out real brain volume change (Takao et al., 2011).

How updates reach speech: Phone changes and speech endpoints ›Software updates and speech readings ›Noise reduction and speech measures ›

Noise or bias

When a change adds noise, and when it adds bias

Participant-varying measurement error can reduce precision. A common additive shift can cancel in an absolute between-arm contrast, while differential errors or unaccounted rescaling can distort the treatment effect on the intended scale.

For continuous endpoints in randomised trials (Nab et al., 2019), random error leaves the treatment effect unbiased but raises the type II error, the chance of missing a real effect. Systematic error, a shift or rescaling, and differential error, which differs by arm, can bias it substantially. A constant shift in both arms leaves the difference intact; a rescaling shrinks or stretches it.

Sources of unevenness

Calendar time

Randomisation balances phones and enrolment dates only on average, so in a concurrent trial later updates reach both arms alike in expectation. Arms enrolled at different times, such as non‑concurrent controls in platform trials, meet the same update at different points of follow‑up. Time trends that differ between arms, or are not additive, can inflate the type 1 error, the rate of false positive findings (Bofill Roig et al., 2022).

Phone mix

Updates reach phone models on different schedules, and phone mix can differ between groups. AUC scores how well something separates two groups; 1 is perfect separation. In Cephalgo's summary of results on the mPower study (doi:10.5281/zenodo.24583552), sustained vowels recorded unsupervised on participants' own iPhones were analysed on 80 held-out participants (40 reporting Parkinson's, 40 controls); a probe using only age, sex, phone model and app version reached AUC 0.61 for Parkinson's versus controls, where 0.5 is chance, so details other than the voice carried some group information. Parkinson's status was self-reported by app users; this is a research result, not a diagnostic or clinical decision tool. These data were contributed by users of the Parkinson mPower mobile application as part of the mPower study developed by Sage Bionetworks and described in Synapse (https://www.synapse.org/mPower). Own‑phone risks ›

Who changes phone

Phones are lost, broken or upgraded, and some participants may move to a carer’s phone as they decline. A treatment that changes decline, side effects or dropout can change this by arm.

Changes that depend on the voice

Processing such as compression, which turns loud sounds down, responds to the signal, so its effect can depend on the voice: in simultaneous recordings of four speakers on phones, laptops and a reference recorder, device differences grew or shrank with the value being measured (Penney et al., 2021). A change can then interact with severity, and with any treatment that changes it.

Slopes and models

How slope and mixed models absorb a step

A linear trend model that omits a measurement step can absorb part of that step into the estimated slope.

A step becomes a slope

For readings spread evenly over follow‑up, a step of size d halfway through fits as a change of 1.5 d over the whole period. With five quarterly visits in a year, a step just before the month‑6 or month‑9 visit fits as 1.2 d, and one just before the month‑3 or month‑12 visit as 0.8 d. This is least‑squares arithmetic.

Mixed models spread it

In a random‑slope model, which fits each participant their own rate of change, participants whose phones changed at different times get different false slopes: slope variance grows, power falls, and the average false slope moves each arm’s rate of decline.

Ratios move when differences hold

Placebo declines 10 points and treatment 7: a 30% slowing. Add a false 3‑point decline to both arms and the difference is still 3 points, but the slowing reads 23%.

A change on a known date can be modelled as a step instead. In simulated platform trials, a step‑function time term fitted across arms kept the type 1 error controlled when time effects were equal across arms and additive on the model scale (Bofill Roig et al., 2022). A version log puts the step on a date.

Reliability

Reliability, ICC and the smallest change you can trust

A processing change is measurement error: it lowers reliability and widens the minimal detectable change, the smallest change in one person that can be told apart from error.

Use an agreement ICC

The intraclass correlation coefficient (ICC) scores how closely repeated readings of the same people agree; 1 is perfect agreement. A consistency ICC ignores an additive offset shared across subjects (Koo and Li, 2016). An absolute-agreement ICC includes that disagreement; participant-specific shifts can affect either form. Report ICC(A,1) with the mean difference.

From ICC to detectable change

Under the usual assumptions (equal error variance, independent errors), the standard error of measurement is SEM = SD × √(1 − ICC), and the 95% minimal detectable change is 1.96 × √2 × SEM (de Vet et al., 2006): about 0.9 standard deviations of the sample at an ICC of 0.9, 1.4 at 0.75 and 2 at 0.5. A systematic step needs its own check.

From ICC to sample size

In the models of Perkins and colleagues, raising reliability from 0.7 to 0.9 cut the required sample size by 22%.

Speech features can start from low reliability

Before any phone change, Stegmann and colleagues found median test‑retest ICCs of 0.02 to 0.55 for open‑source speech and language features, in controls, ALS and suspected frontotemporal dementia.

Estimands and the FDA

Where a phone change sits in the estimand framework

ICH E9(R1) does not address phone or software changes. In our reading, a phone or software change alters the instrument rather than the clinical question, so it is usually handled in the estimator and sensitivity analyses; the FDA guidance asks the analysis plan to define such events.

Estimands and intercurrent events

In the ICH E9(R1) addendum, the estimand is a precise description of the treatment effect a trial sets out to estimate, and the estimator is the method of analysis that computes it. Intercurrent events are events after treatment initiation that affect the interpretation or the existence of the measurements tied to the clinical question, such as discontinuation, additional therapy or death.

A measurement change

What fits is a pre‑specified main estimator that accounts for the change, and sensitivity analyses for the same estimand, such as a tipping‑point analysis, which asks how far an assumption must move before the conclusion changes.

Where the two meet

A phone change can follow the disease, as in a move to a carer’s phone. The FDA cites a comparable event: a cane or walker started after baseline when the technology is an activity tracker.

Analytical validation

Under the DiMe V3 framework, an update that makes new data incompatible with earlier data “could be disastrous for longitudinal studies and meta analyses”, and a changed hardware or software chain calls for new verification or analytical validation studies, or documentation of backward compatibility (Goldsack et al., 2020). The FDA names previously collected data as one way to do this.

“Sponsors should keep a record of the timing and nature of any updates for each DHT.” If the measurement changed, “sponsors should compare data from DHTs before and after the update. Sensitivity analyses may be necessary to evaluate the impact of the update.” They “should specify how these differences will be addressed in the analysis of the trial prior to unblinding”.
U.S. Food and Drug Administration, Digital Health Technologies for Remote Data Acquisition in Clinical Investigations, final guidance, December 2023, section IV.H.4. Section IV.E adds that the analysis plan should define events including “changes in technology during the study”.
What Cephalgo found

What Cephalgo found

In Cephalgo's own checks for its CTAD 2026 poster P446 (Boston, 16–19 November 2026; doi:10.5281/zenodo.24297337), a minor iOS release changed audio processing in measurement mode on both iPhones checked, and Android and web browser updates also changed audio processing settings.

Measurement mode is the iOS audio mode meant to minimise system audio processing (Apple developer documentation). Cephalgo's multi-stream design combines linguistic, acoustic and self-supervised analysis.

People cross a threshold

In the same Cephalgo poster, on three published speech datasets, one simulated audio processing update moved 32% of people to the other group in the hardest comparison, Alzheimer's disease versus MCI in one dataset, with standard handcrafted speech features and 27% with Cephalgo's multi-stream model, although their condition had not changed (doi:10.5281/zenodo.24297337).

Variance from processing

In Cephalgo's ISCTM 2026 poster (Philadelphia, 16 September 2026; doi:10.5281/zenodo.24060676), recordings from the Italian Parkinson's Voice and Speech corpus (28 people with Parkinson's, 22 age-matched controls; 136 recordings from 9 unseen participants) went through three noise-reduction methods standing in for successive phone software versions, and the choice of noise-reduction method explained 2% of the variance in the multi-stream model's output, against 22% for a handcrafted-feature model. Under the change of noise-reduction method, the multi-stream model’s worst-case AUC was 0.927 and its per-recording ICC(A,1) 0.751; the limits are one controlled cohort and a simulated change, and the worst-case specificity of 60.9% rests on 4 control speakers. The recordings come from the Italian Parkinson's Voice and Speech corpus (Dimauro and Girardi, IEEE DataPort, CC BY 4.0); in Cephalgo's posters they were passed through noise-reduction methods, and the dataset's authors have not reviewed these results.

How Cephalgo’s recalibration is designed: each recording is placed in the period in which one processing setting was in force, from the device and software version logged with it or, where none was logged, read from the audio. Recalibration is applied to the measures and runs blind to treatment arm; the audio files are never altered. How Cephalgo recalibrates readings from the audio ›

What to do

What sponsors can do before unblinding

Treat each phone and software change as a dated event, test measures against it, and fix its handling in the analysis plan before unblinding.

  1. Power with that reliability

    Use reliability measured across processing changes in the sample size calculation, and the minimal detectable change for claims about individuals.

  2. Check balance, blinded

    In blinded review, tabulate platforms, updates and replacements by site, country and month; an independent statistician checks them by arm.

  3. Pre‑specify the handling

    Pre-specify how each logged change is handled (a step term, recalibrated values or a covariate) and which sensitivity analysis runs.

  4. Plan the sensitivity analysis

    Add a sensitivity analysis for the same estimand, such as a tipping‑point analysis of how far the handling of changes could move the result.

Logging versions, testing measures and recalibrating from the audio: How do you keep speech readings stable when phones and software change? › Related: Android and iPhone recordings ›BYOD ›Study phones ›

Questions

Phone changes and trial results: common questions

Can a phone or software update bias a clinical trial endpoint?

Yes. An update that reaches both arms alike mostly adds variance and costs power. One that reaches the arms unevenly, rescales the measure or interacts with severity can bias the treatment comparison.

How can a device change look like decline in a longitudinal trial?

A processing change can shift readings in one step, and a slope or mixed model that omits that step can absorb it into the estimated rate of change, as extra decline or improvement. In Cephalgo's simulated year-long trials, one audio processing update made speech timing features decline up to 39% faster or slower than the true decline (CTAD 2026 poster P446, doi:10.5281/zenodo.24297337).

How should the statistical analysis plan handle a phone or software change in a speech endpoint?

Log device model, operating system build, app version and analysis version with every recording; test each measure on the same audio processed different ways; recalibrate readings across processing periods, blind to arm; and fix the handling of changes before unblinding. The full capture, validation and recalibration steps are in How do you keep speech readings stable when phones and software change?

Why do speech biomarker readings shift when phones change during a trial?

The phone's audio processing is part of the measurement. Operating system, app and noise-cancellation updates and replaced phones can change it, so readings can shift in steps that do not come from the participant, a pattern sometimes called drift.

Is a phone or software change an intercurrent event under ICH E9(R1)?

ICH E9(R1) does not address phone or software changes. In our reading, it is usually not one: a phone change alters the instrument rather than the clinical question, so it is handled in the estimator and in sensitivity analyses for the same estimand. The FDA guidance asks the analysis plan to define such events.

Which ICC should be used to test a speech measure across phone changes?

An absolute-agreement ICC, such as ICC(A,1), with the mean difference. A consistency ICC ignores an additive offset shared across subjects, so it misses a step that moves every reading by the same amount.

About Cephalgo

About Cephalgo

Cephalgo SAS, founded in 2021 in Strasbourg, France, builds reliable speech readings for CNS clinical trials in Alzheimer’s disease, Parkinson’s disease, Huntington’s disease and ALS. Cephalgo’s readings are recalibrated from the recorded audio to stay stable when participants’ own phones and software change. Device changes ›Validation approach ›

Funded by the European Union. Cephalgo's work described on this page was supported by the European Innovation Council (EIC) Accelerator (project ADAPTE, Grant Agreement No. 190129251) and by Horizon Europe (grant agreement No 101095436). Views and opinions expressed are however those of the author(s) only and do not necessarily reflect those of the European Union, the European Innovation Council and SMEs Executive Agency (EISMEA) or the European Health and Digital Executive Agency (HaDEA). Neither the European Union nor the granting authorities can be held responsible for them.

Sources

Guidance, statistics and device studies

The guidance, studies and records behind this page, each opened and checked on 27 November 2026.

Guidance and frameworks

Measurement error, reliability and trial design

  • Nab L, Groenwold RHH, Welsing PMJ, van Smeden M. Measurement error in continuous endpoints in randomised trials: problems and solutions. Statistics in Medicine. 2019;38(27):5182–5196. doi:10.1002/sim.8359

    Classical error leaves the treatment effect unbiased but raises type II error; systematic and differential error can bias it substantially.

  • Bofill Roig M, et al. On model‑based time trend adjustments in platform trials with non‑concurrent controls. BMC Medical Research Methodology. 2022;22:228. doi:10.1186/s12874-022-01683-w

    Step‑function time terms control type 1 error when time trends are equal across arms and additive; otherwise the error rate can be inflated.

  • Perkins DO, Wyatt RJ, Bartko JJ. Penny‑wise and pound‑foolish: the impact of measurement error on sample size requirements in clinical trials. Biological Psychiatry. 2000;47(8):762–766. doi:10.1016/S0006-3223(00)00837-4

    Raising reliability from 0.7 to 0.9 reduced the required sample size by 22%.

  • Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. Journal of Chiropractic Medicine. 2016;15(2):155–163. doi:10.1016/j.jcm.2016.02.012

    Explains absolute‑agreement and consistency forms of the ICC; consistency allows a constant offset between measurements.

  • de Vet HCW, Terwee CB, Ostelo RWJG, Beckerman H, Knol DL, Bouter LM. Minimal changes in health status questionnaires: distinction between minimally detectable change and minimally important change. Health and Quality of Life Outcomes. 2006;4:54. doi:10.1186/1477-7525-4-54

    Gives SEM = SD√(1 − R) and MDC = 1.96 × √2 × SEM.

Devices, versions and speech measures

  • Stegmann GM, et al. Repeatability of commonly used speech and language features for clinical applications. Digital Biomarkers. 2020;4(3):109–122. doi:10.1159/000511671

    Median test‑retest ICCs of open‑source speech features were between 0.02 and 0.55 across three datasets.

  • Penney J, Gibson A, Cox F, Proctor M, Szakay A. A comparison of acoustic correlates of voice quality across different recording devices: a cautionary tale. Proc. Interspeech 2021. 2021:1389–1393. doi:10.21437/Interspeech.2021-729

    In simultaneous recordings of four speakers on two phones, two laptops and a reference recorder, many voice quality measures differed by device, and the size of the difference depended on the value measured rather than being a simple offset.

  • Takao H, Hayashi N, Ohtomo K. Effect of scanner in longitudinal studies of brain volume changes. Journal of Magnetic Resonance Imaging. 2011;34(2):438–444. doi:10.1002/jmri.22636

    In 208 adults scanned twice on two scanners of the same model, scanner changes over time could cancel out real brain volume change.

  • Apple. Apple security releases. Accessed 27 November 2026.

    From iOS 18.6.2 (20 August 2025) to iOS 26.6.1 (17 August 2026), fifteen iOS updates reached an iPhone 11 or later kept up to date; some releases cover selected models only. iOS 27 followed on 14 September 2026.

  • Apple. AVAudioSession.Mode.measurement. Apple Developer Documentation. Accessed 27 November 2026.

    The mode for apps that need to minimise system‑supplied signal processing on input and output.

Cephalgo records

  • Noise‑cancellation variability as a source of longitudinal drift in speech‑based digital endpoints for Alzheimer’s disease trials. doi:10.5281/zenodo.24297337CTAD 2026, Boston, 16–19 November 2026, poster P446

  • When the device changes mid‑trial: speech‑based digital endpoint robustness to on‑device preprocessing variability. doi:10.5281/zenodo.24060676ISCTM 2026 Autumn Conference, Philadelphia, 16 September 2026

  • Smartphone speech and the patient‑rated MDS‑UPDRS speech item in Parkinson’s disease: a multi‑stream model evaluated on mPower data. doi:10.5281/zenodo.24583552Summary of results, 27 November 2026

Datasets

  • Dimauro G, Girardi F. Italian Parkinson's Voice and Speech. IEEE DataPort, 2019. doi:10.21227/aw6b-tg17. Licence CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/).

    Recordings of 28 people with Parkinson's and 22 age‑matched controls, used in Cephalgo's ISCTM 2026 poster after noise‑reduction processing; the dataset's authors have not reviewed these results.

  • Bot BM, et al. The mPower study, Parkinson disease mobile data collected using ResearchKit. Scientific Data. 2016;3:160011. doi:10.1038/sdata.2016.11

    These data were contributed by users of the Parkinson mPower mobile application as part of the mPower study developed by Sage Bionetworks and described in Synapse (https://www.synapse.org/mPower).

Has a phone or software change already happened in a trial you are running?

Tell us the length of the trial and the phones it allows, and we will send you the full CTAD 2026 poster P446. A stability audit reads the version logs and the audio to find where the phone or its software changed in each participant's recordings, and recalibrates the readings across those periods.

Request

We use your work email to send you the document and to follow up about it and related work. You can opt out at any time. Privacy policy

Thank you. We will send it to your work email.

Something went wrong. Please write to contact@cephalgo.com.

Thank you! Your submission has been received!
Oops! Something went wrong while submitting the form.